The Cleaning & Storage of Reversed-Phase HPLC Columns

 

Introduction:

Reversed-phase chromatography is by far the most widely used technique in high performance liquid chromatography (HPLC). It is popular because it is applicable to most non-polar analytes and to many ionizable and ionic compounds. Most of the stationary phases used in reversed phase chromatography are hydrophobic in nature. Therefore, analytes are separated by their degree of hydrophobic interaction with the stationary phase and matrices with hydrophobic components can also be retained in a similar manner.

In HPLC analysis, Stationary phase is very important and need to be maintained with more care. Column life will be decided by the maintenance and washing. If the column washed properly after every analysis and stored with end capped, then column life will be increased and the problem raises due to column will get reduce.

What is Reversed-Phase HPLC Columns?

reverse phase column, or reversed-phase HPLC columns, are chromatography columns that contain a non-polar stationary phase.

What Causes Contaminant Build up in Reversed-Phase Columns and How to solve?

Sample matrix:

    Injecting the sample without filtration.

  Usually, sample matrix contains compounds that are of no interest to analysts. If sample matrix components are retained strongly on the column and if the mobile-phase solvent composition itself never becomes strong enough to elute them, these adsorbed or absorbed compounds will accumulate, usually at the head of the column, after many injections.

    Make sure that the sample must be filtered before injecting into the HPLC. Flush the column with appropriate washing solvent at end of every analysis.

pH of the Mobile phase:

    In general, HPLC columns are stable within a pH range of 2 to 8. If you are measuring a pH value, the measurement must be done in the aqueous media before mixing the eluent with organic solvents. Now a days HPLC columns can be used out side of the pH range. Check with the vendor certificate before using the silica-based column outside the pH range of 2 to 8. However, best lifetimes are obtained between pH 2.0 and pH 8.

 Mechanical Vibration:

    Basically, in general Stationary phases based on silica are mechanically very stable. The packed columns show no pressure limit and can be used at more than 40 MPa (6000 psi) without any problem. However, avoid pressure shocks on the column. Pressure shocks lead to channeling in the column, which results in peak splitting in the corresponding chromatogram.

 Mobile Phase & Solvent:

    Silica based stationary phases are compatible with all organic solvents in the above mentioned pH range. Always use correct grade of solvents and salts during the mobile phase preparation.

    It is a very good practice to filter the mobile phase by 0.45µ membrane filter (0.22µ if UPLC) before using in HPLC instrument.

    The use of non pure solvents in HPLC causes irreversible adsorption of impurities on the column head. These impurities block adsorption sites, change the selectivity of the column and lead to peak splitting in the chromatogram. In gradient elution, impurities cause so called “Ghost Peaks”. Ghost peaks are peaks that always appear in the same position on the chromatogram. Their origin is not the sample, but the impurities from the solvents or solvent additives. Therefore, it is highly recommended to run a gradient without injection in the beginning of each method to determine the ghost peaks.

Column Washing & Storage:

    1. A recommended column washing system for a typical bonded-silica column and a mobile phase without buffer salts is to use

    2. After completion of the analysis, flush your column without buffer and store in the proper storage solvent.

    3. For short term storage, like overnight, columns can be stored in the eluent used in last analysis.

   4. For middle term storage, like 2 days or over the weekend, columns should be flushed with pure water to prevent microbial growth.

     5. For long term storage, silica-based columns should be stored in an aprotic solvent. The water content should not be higher than 50%. The best storing solvent is Acetonitrile for most of the solvent. (Storage solvents depends product sometimes)

    6. After the analysis, column should not flush with 100% organic solvent, if buffer is used as mobile. Make sure that all buffers are washed out of the column before flushing with Acetonitrile. Buffer salts are mainly not soluble in Acetonitrile and can block the capillaries and the column.

    7. Do not allow the column to dry out. Keep it tightly capped when not in use.

    8. Use the column in the direction indicated on the label. An unusually high operating pressure may indicate a plugged inlet frit and may be cleared by reversing flow through the column for 10-20 mL.

Conclusion:

     The correct use of an HPLC column is extremely important for the life time of a column and therefore for the benefit of your HPLC analysis.          

Post a Comment

0 Comments