Introduction:
Gas chromatography is one
of the most popular chromatography techniques to separate volatile compounds or
Substances. Here the moving phase is gas such as Nitrogen, helium and
stationary phase is high-boiling liquid that is adsorbed on a solid. There are
many problems arises during GC analysis. GC is a very sensitive instrument and
need to be handle with more care since the gases are used. So, safety is must
in Gas Chromatography analysis.  
Maintaining GC routinely
will prevent many problems. Many
GC problems can be prevented if the column is properly installed and GC is
maintained routinely. For example, replacing septa or liner at regular
intervals and keeping the injector and detector clean and well-maintained
should solve many problems. Regular preventive maintenance depends on
particular model of GC and you should consult required operations.
The starting point of
investigation is depending on the problem, generally need to investigate on the
following,
1.      Gas
Cylinders
2.      Purification
Panel
3.      Gas
lines which connected to instrument
4.      Injector
5.      Column
6.      Detector
The above compartments
should take into the investigation when problem comes.
Commonly facing problems,
- Baseline Problems
- Unusual peak shapes
- Irreproducible peak response
- Peak not eluting
- Split Peaks
- Tailing Peaks
- Shifting of Retention time
Above mentioned problems
are routinely facing issues in the day-to-day analysis. The root cause for
these issues can to identified easily and resolved. Let’s see the possible root
cause and solution for above mentioned problems.
Baseline Problems:
Baseline problems could be divided into 5 categories:
drift, noise, offset, spiking and wander.
·        
Drift means slow baseline movement
in one direction
·        
Noise is rapid and random movement
of the baseline position
·        
Offset is sudden unexplained change
of the baseline position
·        
Spiking is presented by peaks with
no width, either positive or negative
·        
Wander is low frequency noise
| Possible
  Cause | Solution | 
| Drift: Downward
  drift for a few minutes is normal after installing a new column. | Increase
  the oven temperature to close to the maximum continuous operating temperature
  for the column. Maintain the temperature until flat baseline is observed. If
  the detector signal does not drop in 10 minutes, immediately cool the column and
  check for leaks. | 
| Unequilibrated
  detector. | Allow
  sufficient time for temperature equilibration of the detector. | 
| Downward
  drift is frequently due to the "back-out" of contaminants from the
  detector or other parts of the GC. | Clean
  out contamination. | 
| Damage
  to the stationary phase of the GC column. | Determine
  the cause of the damage. It may be due to impurities in the carrier gas or to
  excessive temperatures. Replace column. | 
| Drift
  in gas flow rates. | Clean
  or replace flow or pressure regulator(s). Adjust pressure. | 
| Noise: The
  column may be inserted too far into the flame of an FID. | Reinstall
  the column. Be sure to insert the column into the detector exactly the
  correct distance specified in the instrument manual. | 
| An
  air leak can result in noise in ECD and TCD detectors. | Eliminate
  the leak. | 
| Incorrect
  combustion gases or flow rates can generate noise in FID, NPD or FPD
  detectors. Drift in gas flow rates. | Clean
  injector. Replace inlet liner, septa and seals. | 
| Contaminated
  column. | Bake
  out the column. Cut off first 10 cm of column. If it does not help, replace
  the column. | 
| Defective
  detector. | Clean
  and/or replace parts as necessary. | 
| Defective
  detector board. | Consult
  GC manufacturer. | 
| Spiking: Electrical
  disturbances entering the chromatograph through power cables, even shielded
  cables | Try
  to correlate spikes with events in equipment near the chromatograph.
  Periodicity is often a clue. Turn off equipment or move it. If necessary,
  install a voltage regulator. | 
Unusual peak shapes:
| Possible
  Cause | Solution | 
| Ghost peaks:Remnants of previous samples in the inlet or column.
  Ghost peaks due to remnants are most likely to occur when increasing inlet or
  column temperatures. | Increase
  the final temperature and lengthen the run time to allow for complete elution
  of previous samples. If ghost peaks continue to occur, clean the inlet.
  Condition the column at temperature higher than has been used but lower than
  the maximum continuous operation temperature for the column.  | 
| Bleed
  from the septum or fragments of the septum lodged in the inlet or liner. | Cut
  10 cm off the inlet end of the column and/reverse it before reconditioning
  it. If it does not help, replace the column. Clean
  the inlet. Replace the inlet liner, glass wool and seals. | 
| Fronting peaks:Fronting peaks are usually the result of column
  overloading. In this case, the effect should be a function of injection
  volume. | Solutions
  include reducing the injection volume or using a column with greater
  capacity. Columns with larger diameter or thicker stationary phase coatings
  generally have larger sample capacities, however, resolution may be reduced. | 
Irreproducible peak response:
| Possible
  Cause | Solution | 
| Inconsistent
  injection. | Develop
  a reproducible injection technique. Use autosampler. | 
| Distorted
  peak shapes can adversely affect quantitative determinations. | Correct
  any problems that result in the distortion of peak shape. See Peak shape
  problems. | 
| Variations
  in GC operating parameters. | Standardize
  operating parameters. | 
Peak Not Eluting:
| Possible
  Cause | Solution | 
| Defective
  syringe. | Try
  a new or proven syringe. | 
| "Blown"
  septum or massive leaks at the inlet. | Find
  and fix leaks. | 
| Problems
  with carrier gas flow. | Adjust
  gas flow. Check the column flow and the column outlet by immersion to
  methanol. | 
| Broken
  column or column installed in the wrong way. | Replace
  or reinstall the column. | 
| The
  detector is not functioning or not connected to the recorder or integrator. | Ensure
  that detector is working properly. E.g.: Is the flame in FID on? Check
  connection to the output device. | 
Split Peaks:
| Possible
  Cause | Solution | 
| Fluctuations
  in column temperature. | Repair
  temperature control system | 
| Mixed
  sample solvent for splitless or on-column injections. | Use
  single solvent | 
Tailing Peaks:
| Possible
  Cause | Solution | 
| Contaminated
  or active injector liner, seal or column. | Clean
  or replace injector liner. Do not use glass wool in the liner. If necessary,
  replace the column. | 
| Dead
  volume due to poorly installed liner or column. | Confirm
  by injecting inert peak (methane). If it tails, column is not properly
  installed. Reinstall liner and column as necessary. | 
| A
  bad match between the polarities of the stationary phase and the solvent. | Change
  the stationary phase. Usually polar analytes tail on no-polar columns, or
  dirty columns. | 
| Debris
  in the liner or column. | Clean
  or replace the liner. Cut 10 cm off the end of the column and reinstall it. | 
| Injection
  takes too long. | Improve
  injection technique. | 
| Split
  ration is too low. | Increase
  split ratio to at least 20:1 | 
| Overloading
  the inlet. | Decrease
  the sample volume or dilute sample | 
| Some
  types of compounds such as alcoholic amines, primary and secondary amines and
  carboxylic acids tend to tail. | Try
  a more polar column. Make a derivative of the sample. | 
Shifting of Retention time:
| Possible
  Cause | Solution | 
| Change
  in column temperature. | Check
  GC oven temperature | 
| Change
  in gas flow rate (linear velocity). | Inject
  a detectable unretained sample such as methane to determine the linear gas
  velocity. Adjust gas pressure or flow to obtain proper values for your
  analytical method. | 
| Leak
  in the injector. | Check
  the septum first. Change, if necessary. Find the leak and fix it. | 
| Change
  of solvent. | Use
  the same solvent for standards and samples. | 
| Contaminated
  column. | Bake
  out the column. Cut 10 cm off the end of the column. If necessary, replace
  the column. | 
0 Comments